If the autoclave is not calibrated in accordance with the specified requirements, the parameters are not up to standard, and the operation and maintenance are not performed in accordance with the equipment operating procedures and instructions, the sterilization effect may be reduced or lost. Failure to perform regular autoclave effectiveness evaluation may result in incomplete sterilization due to substandard sterilization effects, which may easily cause blank bacteria to grow, and sample contamination may lead to experimental failure. The temperature in the autoclave is too high, which can easily cause steam burns.
Risk control measures
Set up a dedicated equipment administrator to be responsible for the regular calibration and maintenance of the equipment. Participate in the professional training of autoclave, and hold a certificate.
Establishing procedures for the evaluation of the effects of autoclaves, including monitoring methods and monitoring frequencies. And record relevant data.
Chemical method monitoring refers to putting the sterilization indicator strip into the article to be sterilized. After a sterilization cycle, if the color and appearance meet the specified requirements, it will be qualified, otherwise it will be unqualified.
The indicator tube does not change color (purple) after incubation, indicating that the sterilization has passed. After incubation, the indicator tube changes color (yellow), indicating that the sterilization has not passed. The control tube cultured at the same time should be positive (yellow).
When the pressure in the autoclave drops to zero and the temperature drops to the specified requirements, then take the items. You can wear a high-temperature-proof handbag to avoid burns.
We will sterilize the products, such as gauze bandages, medical absorbent cotton balls, wound dressings, and can provide sterile and non-sterile types.